Volume 37, Issue 12 , Pages 1619-1623, December 2011

Evaluation of Accuracy, Reliability, and Repeatability of Five Dental Pulp Tests

  • Eugene Chen, BDSc, BScDent(Hons), DClinDent
  • ,
  • Paul V. Abbott, BDSc, MDS

      Affiliations

    • Corresponding Author InformationAddress requests for reprints to Professor Paul V. Abbott, School of Dentistry, University of Western Australia, 17 Monash Avenue, Nedlands, WA 6009, Australia.

School of Dentistry, University of Western Australia, Nedlands, Western Australia, Australia

published online 12 September 2011.

Abstract 

Introduction

The aim of this study was to compare the clinical accuracy, reliability, and repeatability of laser Doppler flowmetry (LDF), an electric pulp test (EPT), and various thermal pulp sensibility tests.

Methods

Pulp tests were done on 121 teeth in 20 subjects by using LDF, EPT, and thermal pulp testing (CO2, Endo Frost [EF], Ice) during 2 or 3 test sessions with at least 1-week intervals. The order of testing was reversed on the second visit. A laser Doppler flowmeter was used to measure mean pulp blood flow (Flux) calibrated against a brownian motion medium and zeroed against a static reflector. The laser source was 780 nm, with 0.5-mm fiber separation in the probe, 3.1 kHz as the primary bandwidth for filter set to 0.1-second time output constant. Customized polyvinylsiloxane splints were fabricated for each participant, and a minimum of 90-second recording time was used for each tooth. Raw data were analyzed by using repeated measure analysis of variance, pairwise comparisons, and interclass correlations (ICC).

Results

The accuracy of EPT, CO2, and LDF tests was 97.7%, 97.0%, and 96.3%, respectively, without significant differences (P > .3). Accuracy of EF and Ice was 90.7% and 84.8%, respectively. EPT (P = .015) and CO2 (P = .022) were significantly more accurate than EF. LDF was more accurate than EF, but this was not statistically significant (P = .063). Ice was significantly less accurate than EPT (P = .004), CO2 (P = .005), LDF (P = .006), and EF (P = .019). With the exception of Ice (effect of visit: F2,38 = 5.67, mean squared error = 0.01, P = .007, η2p = 0.23), all tests were reliable. Ice (ICC = 0.677) and LDF (ICC = 0.654) were the most repeatable of the tests, whereas EPT (ICC = 0.434) and CO2 (ICC = 0.432) were less repeatable.

Conclusions

CO2, EPT, and LDF were reliable and the most accurate tests, but CO2 and EPT were less repeatable yet less time-consuming than LDF. EF was reliable but not as accurate as EPT and CO2 and less repeatable than Ice and LDF. Ice was the most repeatable but the least accurate and least reliable test.

Key Words: Electric pulp tests, laser Doppler flowmetry, pulp sensibility tests, thermal pulp tests, vitality tests

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 20.00 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0099-2399(11)00865-X

doi:10.1016/j.joen.2011.07.004

Journal of Endodontics
Volume 37, Issue 12 , Pages 1619-1623, December 2011