Resistance to Flexural Fatigue of Reciproc R25 Files under Continuous Rotation and Reciprocating Movement

Published:January 27, 2012DOI:https://doi.org/10.1016/j.joen.2011.12.033

      Abstract

      Introduction

      The aim of the present work was to evaluate the resistance to flexural fatigue of Reciproc R25 nickel-titanium files, 25 mm, used in continuous rotation motion or reciprocation motion, in dynamic assays device.

      Methods

      Thirty-six Reciproc R25 files were divided into 2 groups (n = 18) according to kinematics applied, continuous rotary (group CR) and reciprocation motion (group RM). The files were submitted to dynamic assays device moved by an electric engine with 300 rpm of speed that permitted the reproduction of pecking motion. The files run on a ring’s groove of temperate steel, simulating instrumentation of a curved root canal with 40° and 5 mm of curvature radius. The fracture of file was detected by sensor of device, and the time was marked. The data were analyzed statistically by Student's t test, with level of significance of 95%.

      Results

      The instruments moved by reciprocating movement reached significantly higher numbers of cycles before fracture (mean, 1787.78 cycles) when compared with instruments moved by continuous rotary (mean, 816.39 cycles).

      Conclusions

      The results showed that the reciprocation motion improves flexural fatigue resistance in nickel-titanium instrument Reciproc R25 when compared with continuous rotation movement.

      Key Words

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Journal of Endodontics
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Yared G.
        Canal preparation using only one Ni-Ti rotary instrument: preliminary observations.
        Int Endod J. 2008; 41: 339-344
        • Sonntag D.
        • Peters O.A.
        Effect of prion decontamination protocols on nickel-titanium rotary surfaces.
        J Endod. 2007; 33: 442-446
        • Alapati S.B.
        • Brantley W.A.
        • Svec T.A.
        • Powers J.M.
        • Mitchell J.C.
        Scanning microscope observations of new and used nickel-titanium rotary files.
        J Endod. 2003; 29: 667-669
        • You S.Y.
        • Bae K.S.
        • Baek S.H.
        • Kum K.Y.
        • Shon W.J.
        • Lee W.
        Lifespan of one nickel-titanium rotary file with reciprocating motion in curved root canals.
        J Endod. 2010; 36: 1991-1994
        • Shen Y.
        • Qian W.
        • Abtin H.
        • Gao Y.
        • Haapasalo M.
        Fatigue testing of controlled memory wire nickel-titanium rotary instruments.
        J Endod. 2011; 37: 997-1001
      1. Yared G. Canal preparation with only one reciprocating instrument without prior hand filing: a new concept. Available at: http://www.vdw-reciproc.de/images/stories/pdf/GY_Artikel_en_WEB.pdf. Accessed December 9, 2011.

        • De-Deus G.
        • Moreira E.J.L.
        • Lopes H.P.
        • Elias C.N.
        Extended cyclic fatigue life of F2 ProTaper instrument used in reciprocating movement.
        Int Endod J. 2010; 43: 1063-1068
        • Varela-Patino P.
        • Ibanez-Parraga A.
        • Rivas-Mundina B.
        • Cantatore G.
        • Otero X.L.
        • Martin-Biedma B.
        Alternating versus continuous rotation: a comparative study of the effect on instrument life.
        J Endod. 2010; 36: 157-159
        • Johnson E.
        • Lloyd A.
        • Kuttler S.
        • Namerow K.
        Comparison between a novel nickel-titanium alloy and 508 nitinol on the cyclic fatigue life of ProFile 25/.04 rotary instruments.
        J Endod. 2008; 34: 1406-1409
        • Gao Y.
        • Shotton V.
        • Wilkinson K.
        • Phillips G.
        • Johnson W.B.
        Effects of raw material and rotational speed on the cyclic fatigue of ProFile Vortex rotary instruments.
        J Endod. 2010; 36: 1205-1209
        • Wu J.
        • Lei G.
        • Yan M.
        • Yu Y.
        • Yu J.
        • Zhang G.
        Instrument separation analysis of multi-used ProTaper Universal rotary system during root canal therapy.
        J Endod. 2011; 37: 758-763
        • Parrashos P.
        • Messer H.H.
        Rotary NiTi instrument fracture and its consequences.
        J Endod. 2006; 32: 1031-1043
        • Pruett J.P.
        • Clement D.J.
        • Carnes Jr., D.L.
        Cyclic fatigue testing of nickel-titanium endodontic instruments.
        J Endod. 1997; 23: 77-85
        • Haïkel Y.
        • Serfaty R.
        • Bateman G.
        • Senger B.
        • Allemann C.
        Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments.
        J Endod. 1999; 25: 434-440
        • Yared G.M.
        • Bou Dagher F.E.
        • Machtou P.
        Cyclic fatigue of ProFile rotary instruments after clinical use.
        Int Endod J. 2000; 33: 204-207
        • Li U.M.
        • Lee B.S.
        • Shih C.T.
        • Lan W.H.
        • Lin C.P.
        Cyclic fatigue of endodontic nickel-titanium rotary instruments: static and dynamic tests.
        J Endod. 2002; 28: 448-451
        • Yao J.H.
        • Schwartz S.A.
        • Beeson T.J.
        Cyclic fatigue of three types of rotary nickel-titanium files in a dynamic model.
        J Endod. 2006; 32: 55-57
        • Ray J.J.
        • Kirkpatrick T.C.
        • Rutledge R.E.
        Cyclic fatigue of EndoSequence and K3 rotary files in a dynamic model.
        J Endod. 2007; 33: 1469-1472
        • Gambarini G.
        • Grande N.M.
        • Plotino G.
        • et al.
        Fatigue resistance of engine-driven rotary nickel titanium instruments produced by new manufacturing methods.
        J Endod. 2008; 34: 1003-1005
        • Spanaki-Voreadi A.P.
        • Kerezoudis N.P.
        • Zinelis S.
        Failure mechanism of ProTaper Ni-Ti rotary instruments during clinical use: fractographic analysis.
        Int Endod J. 2006; 39: 171-178
        • Galvão Barbosa F.O.
        • Ponciano Gomes J.A.
        • Pimenta de Araújo M.C.
        Influence of previous angular deformation on flexural fatigue resistance of K3 nickel-titanium rotary instruments.
        J Endod. 2007; 33: 1477-1480
        • Zelada G.
        • Varela P.
        • Martín B.
        • Bahillo J.G.
        • Magán F.
        • Ahn S.
        The effect of rotational speed and the curvature of root canals on the breakage of rotary endodontic instruments.
        J Endod. 2002; 28: 540-542
        • Yared G.M.
        • Bou Dagher F.E.
        • Machtou P.
        Influence of rotational speed, torque and operator’s proficiency on ProFile failures.
        Int Endod J. 2001; 34: 47-53
        • Mesgouez C.
        • Rilliard F.
        • Matossian L.
        • Nassiri K.
        • Mandel E.
        Influence of operator experience on canal preparation time when using the rotary Ni-Ti ProFile system in simulated curved canals.
        Int Endod J. 2003; 36: 161-165
        • Gavini G.
        • Pessoa O.F.
        • Barletta F.B.
        • Vasconcellos M.A.Z.
        • Caldeira C.L.
        Cyclic fatigue resistance of rotary nickel-titanium instruments submitted to nitrogen ion implantation.
        J Endod. 2010; 36: 1183-1186
        • Melo M.C.C.
        • Bahia M.G.A.
        • Buono V.T.L.
        Fatigue resistance of engine-driven rotary nickel-titanium endodontic instruments.
        J Endod. 2002; 28: 765-769
        • Christ H.J.
        Fundamental mechanisms of fatigue and fracture.
        Stud Health Technol Inform. 2008; 133: 56-67
        • Askeland D.
        • Phule P.
        The science of engineering of materials.
        4th ed. Brooks/Cole-Thompson Learning, Pacific Grove, CA2003
        • Vaidyanathan R.
        • Dunand D.C.
        • Ramamurty U.
        Fatigue crack-growth in shape-memory NiTi and NiTi-TiC composites.
        Mater Sci Eng A. 2000; 289: 208-216