Advertisement

Minimal Detectable Width of Tooth Fractures Using Magnetic Resonance Imaging and Method to Measure

Published:September 10, 2022DOI:https://doi.org/10.1016/j.joen.2022.08.011

      Abstract

      Introduction

      Vertical root fracture (VRF) in root canal–treated (RCT) teeth is a common cause of pain, bone resorption, and tooth loss. VRF is also difficult to diagnose and measure. Magnetic resonance imaging (MRI) has the potential to identify VRF due to beneficial partial volume averaging, without using ionizing radiation. This investigation aimed to describe the narrowest VRFs detectable based on MRI, using micro-computed tomography (microCT) as the reference standard and proposes a method using profile integrals to measure the widths of small VRFs.

      Methods

      VRFs were induced in 62 RCT tooth root samples. All samples were imaged in a phantom using MRI and reference imaging was obtained using microCT. The stacks of 3-dimensional axial MRIs were assessed by 3 board-certified endodontists. Evaluators determined the most coronal slice within the stack that was discernible as the extent of the VRF. This slice was measured on correlated microCT sections to determine the minimum VRF width (μm) detectable using a profile integral-based method to measure small fractures and negate the effects of the point spread function.

      Results

      Using profile integrals to measure VRF width was repeatable and resulted in estimates that were on average 1 μm smaller than known reference widths. Adjusted median VRF width detected using MRI was 45 μm (first quartile: 26 μm, third quartile: 64 μm).

      Conclusion

      Using profile integrals is a valid way to estimate small VRF width. The MRI approach demonstrated ability to repeatedly detect VRFs as small as 26 μm.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Journal of Endodontics
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • American Association of Endodontists
        Cracking the cracked tooth code: detection and treatment of various longitudinal tooth fractures.
        in: Endodontics Colleagues for Excellence. 2008
        • Weine F.S.
        • Wax A.H.
        • Wenckus C.S.
        Retrospective study of tapered, smooth post systems in place for 10 years or more.
        J Endod. 1991; 17: 293-297
        • Torbjörner A.
        • Karlsson S.
        • Odman P.A.
        Survival rate and failure characteristics for two post designs.
        J Prosthet Dent. 1995; 73: 439-444
        • Morfis A.S.
        Vertical root fractures.
        Oral Surg Oral Med Oral Pathol. 1990; 69: 631-635
        • Vire D.E.
        Failure of endodontically treated teeth: classification and evaluation.
        J Endod. 1991; 17: 338-342
        • Zadik Y.
        • Sandler V.
        • Bechor R.
        • Salehrabi R.
        Analysis of factors related to extraction of endodontically treated teeth.
        Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008; 106: e31-e35
        • Touré B.
        • Faye B.
        • Kane A.W.
        • et al.
        Analysis of reasons for extraction of endodontically treated teeth: a prospective study.
        J Endod. 2011; 37: 1512-1515
        • Fuss Z.
        • Lustig J.
        • Tamse A.
        Prevalence of vertical root fractures in extracted endodontically treated teeth.
        Int Endod J. 1999; 32: 283-286
        • Baageel T.M.
        • Allah E.H.
        • Bakalka G.T.
        • et al.
        Vertical root fracture: biological effects and accuracy of diagnostic imaging methods.
        J Int So Prev Community Dent. 2016; 6: S93-S104
        • Bernardes R.A.
        • de Moraes I.G.
        • Hungaro Duarte M.A.
        • et al.
        Use of cone-beam volumetric tomography in the diagnosis of root fractures.
        Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009; 108: 270-277
        • Wang S.
        • Xu Y.
        • Shen Z.
        • et al.
        The extent of the crack on artificial simulation models with CBCT and periapical radiography.
        PLoS One. 2017; 12: e0169150
        • Valizadeh S.
        • Vasegh Z.
        • Rezapanah S.
        • et al.
        Effect of object position in cone beam computed tomography field of view for detection of root fractures in teeth with intra-canal posts.
        Iran J Radiol. 2015; 12: e25272
        • Moudi E.
        • Haghanifar S.
        • Madani Z.
        • et al.
        Assessment of vertical root fracture using cone-beam computed tomography.
        Imaging Sci Dent. 2014; 44: 37-41
        • Hassan B.
        • Metska M.E.
        • Ozok A.R.
        • et al.
        Comparison of five cone beam computed tomography systems for the detection of vertical root fractures.
        J Endod. 2010; 36: 126-129
        • Patel S.
        • Brady E.
        • Wilson R.
        • et al.
        The detection of vertical root fractures in root filled teeth with periapical radiographs and CBCT scans.
        Int Endod J. 2013; 46: 1140-1152
        • Zhang L.
        • Wang T.
        • Cao Y.
        • et al.
        In vivo detection of subtle vertical root fracture in endodontically treated teeth by cone-beam computed tomography.
        J Endod. 2019; 45: 856-862
        • Chavda R.
        • Mannocci F.
        • Andiappan M.
        • Patel S.
        Comparing the in vivo diagnostic accuracy of digital periapical radiography with cone-beam computed tomography for the detection of vertical root fracture.
        J Endod. 2014; 40: 1524-1529
        • Walton R.E.
        Vertical root fracture: factors related to identification.
        J Am Dent Assoc. 2017; 148: 100-105
        • PradeepKumar A.R.
        • Shemesh H.
        • Jothilatha S.
        • et al.
        Diagnosis of vertical root fractures in restored endodontically treated teeth: a time-dependent retrospective cohort study.
        J Endod. 2016; 42: 1175-1180
        • Upton A.C.
        The state of the art in the 1990's: NCRP Report No. 136 on the scientific bases for linearity in the dose-response relationship for ionizing radiation.
        Health Phys. 2003; 85: 15-22
        • Schuurmans T.J.
        • Nixdorf D.R.
        • Idiyatullin D.S.
        • et al.
        Accuracy and reliability of root crack and fracture detection in teeth using magnetic resonance imaging.
        J Endod. 2019; 45: 750-755.e2
        • Idiyatullin D.
        • Garwood M.
        • Gaalaas L.
        • Nixdorf D.R.
        Role of MRI for detecting micro cracks in teeth.
        Dentomaxillofac Radiol. 2016; 45: 20160150
        • Di Nardo D.
        • Gambarini G.
        • Capuani S.
        • Testarelli L.
        Nuclear magnetic resonance imaging in endodontics: a review.
        J Endod. 2018; 44: 536-542
        • Molen A.D.
        Considerations in the use of cone-beam computed tomography for buccal bone measurements.
        Am J Orthod Dentofacial Orthop. 2010; 137: S130-S135
        • Ozer S.Y.
        Detection of vertical root fractures of different thicknesses in endodontically enlarged teeth by cone beam computed tomography versus digital radiography.
        J Endod. 2010; 36: 1245-1249
        • Kapralos V.
        • Koutroulis A.
        • Irinakis E.
        • et al.
        Digital subtraction radiography in detection of vertical root fractures: accuracy evaluation for root canal filling, fracture orientation and width variables. An ex-vivo study.
        Clin Oral Investig. 2020; 24: 3671-3681
        • Huang C.C.
        • Chang Y.C.
        • Chuang M.C.
        • et al.
        Analysis of the width of vertical root fracture in endodontically treated teeth by 2 micro-computed tomography systems.
        J Endod. 2014; 40: 698-702
        • Idiyatullin D.
        • Corum C.A.
        • Nixdorf D.R.
        • Garwood M.
        Intraoral approach for imaging teeth using the transverse B1 field components of an occlusally oriented loop coil.
        Magn Reson Med. 2014; 72: 160-165
        • Idiyatullin D.
        • Corum C.A.
        • Garwood M.
        Multi-band-SWIFT.
        J Magn Reson. 2015; 251: 19-25
        • Brady E.
        • Mannocci F.
        • Brown J.
        • et al.
        A comparison of cone beam computed tomography and periapical radiography for the detection of vertical root fractures in nonendodontically treated teeth.
        Int Endod J. 2014; 47: 735-746