Advertisement

Effect of Periodic Changes in Rotation Speed on Torsional Stress and Screw-in Force by Alternative Rotation Technique

Published:November 11, 2022DOI:https://doi.org/10.1016/j.joen.2022.11.003

      Abstract

      Introduction

      This study evaluated the effect of periodic changes in rotation speed on torsional stress and screw-in force using the dedicated alternative rotation technique (ART) motion of the EQ-M (Metabiomed, Cheongju, Korea) endodontic motor.

      Methods

      Two ART modes of the EQ-M motor in 2 alternative techniques (ART30 and ART50) and continuous rotation were compared using ProTaper Next X2 (Dentsply Sirona, Charlotte, NC) files and simulated resin blocks (n = 12 per group). ART30 and ART50 were operated by continuous rotation of 350 rpm for 360° and then rotated at 30% increased speed from the base speed for 180° and at 50% increased speed for 180°, respectively. Before the test, the simulated resin blocks were pre-enlarged using ProTaper Gold S1 and S2 (Dentsply Sirona) and fixed on a metal stage connected to the force- and torque-measuring unit. During shaping the simulated canal in an automatic up-and-down manner, the parameters of maximum torque, sum of torque, maximum screw-in force, and maximum apical force were measured. The data were statistically analyzed using 1-way analysis of variance and the Tukey post hoc comparison test at the 95% significance level.

      Results

      The ART30 and ART50 groups showed a lower maximum torque, sum of torques, screw-in force, and apical drive force than the continuous rotation group. There was no significant difference between the ART30 and ART50 groups (P > .05). When the screw-in force increased suddenly, the torque correspondently increased.

      Conclusions

      Under the limitations of this study, the ART mode could reduce the torsional stress and apical forces of the screw-in during instrumentation in comparison with continuous rotation.

      Key Words

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Journal of Endodontics
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Sonntag D.
        • Guntermann A.
        • Kim S.K.
        • Stachniss V.
        Root canal shaping with manual stainless steel files and rotary Ni-Ti files performed by students.
        Int Endod J. 2003; 36: 246-255
        • Walia H.
        • Brantley W.A.
        • Gerstein H.
        An initial investigation of the bending and torsional properties of nitinol root-canal files.
        J Endod. 1988; 14: 346-351
        • Sattapan B.
        • Nervo G.J.
        • Palamara J.E.
        • et al.
        Defects in rotary nickel-titanium files after clinical use.
        J Endod. 2000; 26: 161-165
        • Diemer F.
        • Calas P.
        Effect of pitch length on the behavior of rotary triple helix root canal instruments.
        J Endod. 2004; 30: 716-718
        • Ha J.H.
        • Park S.S.
        Influence of glide path on the screw-in effect and torque of nickel-titanium rotary files in simulated resin root canals.
        Restor Dent Endod. 2012; 37: 215-219
        • Ha J.H.
        • Kwak S.W.
        • Kim S.K.
        • et al.
        Screw-in forces during instrumentation by various file systems.
        Restor Dent Endod. 2016; 41: 304-309
        • Adorno C.G.
        • Yoshioka T.
        • Suda H.
        Crack initiation on the apical root surface caused by three different nickel-titanium rotary files at different working lengths.
        J Endod. 2011; 37: 522-525
        • Peters O.A.
        Current challenges and concepts in the preparation of root canal systems: a review.
        J Endod. 2004; 30: 559-567
        • Ha J.H.
        • Kim S.K.
        • Cohenca N.
        • et al.
        Effect of R-phase heat treatment on torsional resistance and cyclic fatigue fracture.
        J Endod. 2013; 39: 389-393
        • Kim T.O.
        • Cheung G.S.
        • Lee J.M.
        • et al.
        Stress distribution of three NiTi rotary files under bending and torsional conditions using a mathematic analysis.
        Int Endod J. 2009; 42: 14-21
        • Shen Y.
        • Zhou H.M.
        • Wang Z.
        • et al.
        Phase transformation behavior and mechanical properties of thermomechanically treated K3XF nickel-titanium instruments.
        J Endod. 2013; 39: 919-923
        • Gambarini G.
        • Plotino G.
        • Grande N.M.
        • et al.
        Mechanical properties of nickel-titanium rotary instruments produced with a new manufacturing technique.
        Int Endod J. 2011; 44: 337-341
        • Thompson S.A.
        • Dummer P.M.
        Shaping ability of ProFile.04 Taper Series 29 rotary nickel-titanium instruments in simulated root canals. Part 1.
        Int Endod J. 1997; 30: 1-7
        • Thompson S.A.
        • Dummer P.M.
        Shaping ability of ProFile.04 Taper Series 29 rotary nickel-titanium instruments in simulated root canals. Part 2.
        Int Endod J. 1997; 30: 8-15
        • West J.
        Progressive taper technology: rationale and clinical technique for the new ProTaper universal system.
        Dent Today. 2006; 25: 66-69
        • Hieawy A.
        • Haapasalo M.
        • Zhou H.
        • et al.
        Phase transformation behavior and resistance to bending and cyclic fatigue of ProTaper gold and ProTaper universal instruments.
        J Endod. 2015; 41: 1134-1138
        • Pereira E.S.
        • Peixoto I.F.
        • Viana A.C.
        • et al.
        Physical and mechanical properties of a thermomechanically treated NiTi wire used in the manufacture of rotary endodontic instruments.
        Int Endod J. 2012; 45: 469-474
        • Ataya M.
        • Ha J.H.
        • Kwak S.W.
        • et al.
        Mechanical properties of orifice preflaring nickel-titanium rotary instrument heat treated using T-wire technology.
        J Endod. 2018; 44: 1867-1871
        • Kimura S.
        • Ebihara A.
        • Maki K.
        • et al.
        Effect of optimum torque reverse motion on torque and force generation during root canal instrumentation with crown-down and single-length techniques.
        J Endod. 2020; 46: 232-237
        • Yared G.
        Canal preparation using only one Ni-Ti rotary instrument: preliminary observations.
        Int Endod J. 2008; 41: 339-344
        • You S.Y.
        • Bae K.S.
        • Baek S.H.
        • et al.
        Lifespan of one nickel-titanium rotary file with reciprocating motion in curved root canals.
        J Endod. 2010; 36: 1991-1994
        • Jamleh A.
        • Alfouzan K.
        Vertical load induced with twisted file adaptive system during canal shaping.
        J Endod. 2016; 42: 1811-1814
        • Schneider S.W.
        A comparison of canal preparations in straight and curved root canals.
        Oral Surg Oral Med Oral Pathol. 1971; 32: 271-275
        • Ha J.H.
        • Cheung G.S.
        • Versluis A.
        • et al.
        'Screw-in' tendency of rotary nickel-titanium files due to design geometry.
        Int Endod J. 2015; 48: 666-672
        • Burklein S.
        • Stuber J.P.
        • Schafer E.
        Real-time dynamic torque values and axial forces during preparation of straight root canals using three different endodontic motors and hand preparation.
        Int Endod J. 2019; 52: 94-104
        • Maki K.
        • Ebihara A.
        • Kimura S.
        • et al.
        Effect of different speeds of up-and-down motion on canal centering ability and vertical force and torque generation of nickel-titanium rotary instruments.
        J Endod. 2019; 45: 68-72
        • Tokita D.
        • Ebihara A.
        • Nishijo M.
        • et al.
        Dynamic torque and vertical force analysis during nickel-titanium rotary root canal preparation with different modes of reciprocal rotation.
        J Endod. 2017; 43: 1706-1710
        • Peters O.A.
        • Paque F.
        Root canal preparation of maxillary molars with the self-adjusting file: a micro-computed tomography study.
        J Endod. 2011; 37: 53-57
        • Shemesh H.
        • Bier C.A.
        • Wu M.K.
        • et al.
        The effects of canal preparation and filling on the incidence of dentinal defects.
        Int Endod J. 2009; 42: 208-213
        • Versiani M.A.
        • Leoni G.B.
        • Steier L.
        • et al.
        Micro-computed tomography study of oval-shaped canals prepared with the self-adjusting file, Reciproc, WaveOne, and ProTaper universal systems.
        J Endod. 2013; 39: 1060-1066
        • Zuolo M.L.
        • De-Deus G.
        • Belladonna F.G.
        • et al.
        Micro-computed tomography assessment of dentinal micro-cracks after root canal preparation with TRUShape and self-adjusting file systems.
        J Endod. 2017; 43: 619-622
        • Lim K.C.
        • Webber J.
        The validity of simulated root canals for the investigation of the prepared root-canal shape.
        Int Endod J. 1985; 18: 240-246
        • Kwak S.W.
        • Ha J.H.
        • Cheung G.S.
        • et al.
        Comparison of In Vitro torque generation during instrumentation with adaptive versus continuous movement.
        J Endod. 2019; 45: 803-807
        • Kyaw M.S.
        • Ebihara A.
        • Kasuga Y.
        • et al.
        Influence of rotational speed on torque/force generation and shaping ability during root canal instrumentation of extracted teeth with continuous rotation and optimum torque reverse motion.
        Int Endod J. 2021; 54: 1614-1622
        • Ha J.H.
        • Kwak S.W.
        • Kim S.K.
        • et al.
        Effect from rotational speed on torsional resistance of the nickel-titanium instruments.
        J Endod. 2017; 43: 443-446
        • Gao Y.
        • Shotton V.
        • Wilkinson K.
        • et al.
        Effects of raw material and rotational speed on the cyclic fatigue of ProFile vortex rotary instruments.
        J Endod. 2010; 36: 1205-1209